ISSN print edition: 0366-6352
ISSN electronic edition: 1336-9075
Registr. No.: MK SR 9/7
Published monthly
|
Single-Purpose Atomic Absorption Spectrometer AMA-254 for Mercury Determination and its Performance in Analysis of Agricultural and Environmental Materials
J. Száková, D. Kolihová, D. Miholová, and P. Mader
Faculty of Agronomy, Czech University of Agriculture, CZ-165 21 Prague, Czech Republic
E-mail: szakova@af.czu.cz
Received: 30 June 2003
Abstract:
A single-purpose atomic absorption spectrometer based on in situ dry ashing followed by gold
amalgamation cold vapour AAS was evaluated. Various agricultural and environmental certified
reference materials (CRMs) and internal reference materials of wide range of mercury contents
were analyzed and good agreement with certified values was found. Control chart of the results of
mercury determination in CRM 12-02-03 (Lucerne) as well as the results of interlaboratory trials
suggested that the system of mercury determination was under control. Evidently, wide spectrum
of agricultural and environmental materials can be analyzed for mercury contents even at levels less
than 1 μg kg−1.
Full paper in Portable Document Format: 585a311.pdf
Chemical Papers 58 (5) 311–315 (2004)
|