ISSN print edition: 0366-6352
ISSN electronic edition: 1336-9075
Registr. No.: MK SR 9/7
Published monthly
|
PH* values of the standard succinate buffer solution in the 50 mass % ethanol-water solvent
E. Kozáková, B. Cséfalvayová, and J. Zajacová
Department of Analytical Chemistry, Faculty of Natural Sciences,
Komenský University, 842 15Bratislava
Abstract: The voltage of a cell without liquid junction consisting of hydrogen
electrode, silver/silver chloride reference electrode, and standard succinate
buffer solution in the 50 mass % ethanol—water solvent was measured
potentiometrically. The measured values of EMF were used for calculating the
conventional paH. values of the succinate buffer solution. By means of these
values the calibration of pH-meter was performed. Furthermore, different
buffer solutions and standards in the 50 mass % ethanol—water solvent were
used as electrolytes in some types of Potentiometrie cells the voltage of which
was measured in the working pH* scale. The calibration of the measured cell
with respect to standard in the mixed solvent was compared with the calibration
with respect to the pH standard in water. The pH* values from the working
scales were confronted with the pH* values obtained from the concentration
pH* scale.
Full paper in Portable Document Format: 363a345.pdf
Chemical Papers 36 (3) 345–351 (1982)
|