ISSN print edition: 0366-6352
ISSN electronic edition: 1336-9075
Registr. No.: MK SR 9/7
Published monthly
|
Determination of cobalt(II) by controlled potential coulometry
M. Čakrt, J. Berčík, and Z. Hladký
Department of Analytical Chemistry, Slovak Technical University,
880 37 Bratislava
Abstract: The determination of Co(II) in the presence of Fe(II), Ni(II), Mn(II),
Cr(III), and Zn(II) by controlled potential coulometry is described. The method
is based on the oxidation of tris(l,10-phenanthroline) cobalt(II) complex
at +0.40 - + 0.50 V vs. saturated silver chloride electrode at platinum
and glassy carbon electrodes in a supporting electrolyte containing an excess
of the complexing agent at pH 3.0 — 4.8. Fe(III) causes interference. The
determination of the quantity of electricity passed using the log i vs. t plots
eliminates the need foi an integrator. The estimate standard deviation for the
determination of 2.0 X 10-8 g-ion Co(II) is 1.6%.
Full paper in Portable Document Format: 274a446.pdf
Chemical Papers 27 (4) 446–450 (1973)
|